Get Permission Pawar, Katkade, Mahale, Khalikar, and Rajguru: Comparative evaluation of accuracy of post space scanned with three different techniques


Introduction

The main aim of prosthodontist is to replace missing teeth, provide and maintain aesthetics and function to remaining teeth. Root canal treatment removes the pulp from tooth making tooth pulp less.1, 2, 3 Thereby making the tooth non vital and significantly it leads to moisture loss from tooth. This makes to tooth brittle which so it becomes susceptible to fracture under any stress. To avoid such circumstances post procedures are recommended to reconstruct the tooth with inadequate coronal portion.4 According to GPT 9- Post & core defined as a post with incorporated core; it provides retention and resistance for an artificial crown; it is also used as a platform for retentive attachment systems and non-retentive overdenture abutments.5

The long-term survival of post depends upon its fit and precision of impression.6 Trueness & Precision both the elements comprises of Accuracy.7 Various endodontic treatments such as incomplete root canal treatment, file breakage, instrument separation, ledge formation affects the survival of post to the post space preparation.6

In fabricating customized post restorations, silicone impression technique is used for fabrication procedures conventionally.8 Conventional impression materials are frequently used in general dental practice and have good properties, but they show some limitations in this respect.9, 10, 11, 12 To eliminate these disadvantages, intraoral digital systems have risen and are becoming popular day by day providing digital workflow.13, 14

Aim

The aim of this in vitro study is to compare three different techniques of scanning post space at different depths. The first technique involves the Post space scanning with an intraoral scanner. The second technique involves scanning the impression of post space and third involves scanning the cast with extraoral scanner.

Materials and Methods

  1. Fifteen Freshly extracted Single rooted teeth were selected and endodontically treated. Obturation was done with conventional methods.

  2. Teeth were then embedded in acrylic models.

  3. Post space were created of the length 9mm. A 4 no (1.30) peeso reamer was used up to 3 mm, 3 no (1.10) peeso reamer up to 6 mm & 2 no (0.90) peeso reamer up to 9mm.

  4. Total Sample size taken was fifteen. There were three groups each with sample size of five.

  5. First group consist of post space scanned with Omnicam intra oral scanner. Second group consist of Post space pattern scanned with InEos X5 Extraoral scanner. Third group consist of Post space impression made with Addition silicone, Poured and scanned with InEos X5 Extraoral scanner.

  6. Scanning files of all the three techniques were then converted to standard STL files.

  7. Files were opened in autodesk mesh mixer software where the length and diameter of post space measured.

  8. Diameter was measured at three different depths i.e., 3mm,6mm & 9mm.

  9. Also, height & diameter was measured by 2 examiners to reduce bias.

Table 1

Height & diameter of post spaceat different depths using three different scanning techniques examined by 2 examiners

Height

Diameter

3 mm

6 mm

9 mm

E1

E2

E1

E2

E1

E2

E1

E2

Intra Oral

1

8.91

8.89

1.31

1.32

1.01

1.02

0.80

0.81

2

8.92

8.93

1.30

1.29

1.05

1.03

0.79

0.78

3

8.88

8.87

1.28

1.30

1.00

1.01

0.79

0.80

4

8.90

8.91

1.33

1.30

1.04

1.03

0.81

0.82

5

8.85

8.88

1.30

1.31

1.02

1.05

0.78

0.77

Extra Oral

1

8.40

8.43

1.29

1.28

0.99

0.98

0.76

0.73

2

8.45

8.41

1.28

1.28

1.00

1.01

0.77

0.76

3

8.42

8.29

1.30

1.29

0.98

0.99

0.77

0.75

4

8.39

8.36

1.27

1.25

1.02

1.00

0.75

0.77

5

8.44

8.45

1.29

1.26

1.01

1.02

0.78

0.79

CAST

1

8.11

8.01

1.25

1.23

0.98

0.97

0.71

0.72

2

7.97

7.95

1.24

1.25

0.96

0.98

0.72

0.73

3

8.02

7.98

1.24

1.26

1.00

0.99

0.69

0.70

4

7.99

8.10

1.26

1.24

0.99

1.01

0.70

0.71

5

8.00

8.21

1.27

1.25

0.98

0.96

0.71

0.72

[i] E1: Examiner 1, E2: Examiner 2

Figure 1

Intraoral scanner dentsply CEREC omnicam

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/4c08b6cb-fadc-44b4-8831-568ce77dc932/image/e7ceb53b-97fc-4ef6-bb78-703ada66470f-uimage.png

Figure 2

Extraoral scanner

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/4c08b6cb-fadc-44b4-8831-568ce77dc932/image/4b06f0af-ea95-4ccd-9616-c51654f45f76-uimage.png

Figure 3

Autodesk mesh mixer

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/4c08b6cb-fadc-44b4-8831-568ce77dc932/image/43193cd0-ae81-4964-8872-d407d4080e7c-uimage.png

Result

One-way ANOVA and Tukey’s post hoc test was used to know comparison. Group 1 showed mean height of 8.89 mm. Group 2 showed mean height of 8.4 mm. While group 3 showed mean height of 8.03 mm. So, the intraoral scanner showed the best accuracy in measuring the height.

  1. Group 1 showed mean diameter of 1.3 mm at 3mm. Group 2 showed mean diameter of 1.27 mm at 3mm. Group 3 showed mean diameter of 1.24 mm at 3mm.

  2. Group 1 showed mean diameter of 1.02 mm at 6mm. Group 2 showed mean diameter of 1 mm at 6mm. Group 3 showed mean diameter of 0.98 mm at 6mm.

  3. Group 1 showed mean diameter of 0.795 mm at 9mm. Group 2 showed mean diameter of 0.763 mm at 9mm.Group 3 showed mean diameter of 0.711 mm at 9mm. So, the intra oral scanner showed the best accuracy of diameter and as the depth increases the accuracy decreases.

Graph 1
https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/4c08b6cb-fadc-44b4-8831-568ce77dc932/image/bb78317a-b8de-44f1-b84f-cca448063b6c-uimage.png

Graph 2
https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/4c08b6cb-fadc-44b4-8831-568ce77dc932/image/ee5910d9-427f-4288-aa4e-33830ab319d6-uimage.png

Discussion

The aim of this study was to evaluate the accuracy of scanning post space using three different techniques at different depths. In this research the probability of creating the post scan with the use of three different scanning techniques have been examined. Advancements in scanners and different techniques will ultimately lead to the road for a more precise scans and more applications will allow for future research to assess their accuracy and precision.7, 15

The use of computer-aided manufacturing and computer aided design is increasing for its applications in dentistry. CAD/CAM production systems provides variety of laboratory procedures and overcome the drawbacks of conventional methods. However, manufacturers can pay enough attention for the development of CAD/CAM technology for milling of custom-made post and cores. The CAD-CAM design of post and cores allows precise planning of the clinical treatment and accurate fabrication of the post and core, thereby reducing treatment expenditure.

Limitations

This study has some limitations, as

  1. It was conducted in vitro & did not mimic oral conditions.

  2. The effects of temperature and humidity of the oral environments, saliva and blood, soft tissues, patient movement, oral cavity, and scanning laser angle of incidence were not considered.

  3. At the same time, many items in oral cavity, such as implants (ceramics, metals & composite resin), dentin enamel, oral soft tissue and different materials and geometric shapes, also affect the accuracy of scanning.

But with the advent of technology these limitations can be overcome in future.

Conclusion

Intraoral scanners showed the better accuracy and precision for scanning the post space as compared to extra oral scanner. Also, greater depths of the post space preparation showed varying degrees of trueness. As the depth of post space increases, the precision and accuracy decreases. Intraoral scanners can be used to obtain post space impression data as an alternative to conventional silicone impression method. This system has good and reliable results in reading post space preparation.

Source of Funding

None.

Conflict of Interest

None.

References

1 

M Benli B Gokcen-Rohlig Use of intraoral scanning device for the impression of post space: an in-vitro studyInt J Curr Res20181010747657

2 

M Ferrari MC Cagidiaco S Grandini M DeSanctis C Goracci Post placement affects survival of endodontically treated premolarsJ Dent Res200786872934

3 

P Schmage P Pfeiffer E Pinto U Platzer I Nergiz Influence of oversized dowel space preparation on the bond strengths of FRC postsOper Dent200934193101

4 

H Kalra U Sukhija RR Rassawet V Rani Dr Rani A Review on Post and CoreSch J Dent Sci20207351710.36347/sjds.2020.v07i03.002

5 

S Uniyal H Aeran B Kwatra A Nautiyal Post & Core: An easy and effective treatment modality for severely damaged teethInt J Oral Health Dent20151299104

6 

AA Hegazi MM Zohdy TS Morsi Effect of Two Post Space Depths on the Accuracy of the ScanMansoura J Dent2022912731

7 

The Glossary of Prosthodontic Terms: Ninth EditionJ Prosthet Dent20171175S1105

8 

M Sacher G Schulz H Deyhle K Jäger B Müller B Müller G Wang Comparing the accuracy of intraoral scanners, using advanced micro computed tomographyDevelopments in X-Ray Tomography XII92019SPIE59

9 

T Nino J Juloski C Michele C Goracci V Allesandro S Grandini Influence of different scanning techniques on in vitro performance of CAD-CAM fabricated fiber postsJ Oral Sci20186022628

10 

AB Nulty A comparison of full arch trueness and precision of nine intra-oral digital scanners and four lab digital scannersDent J (Basel)20219775

11 

A Ender A Mehl In-vitro evaluation of the accuracy of conventional and digital methods of obtaining full-arch dental impressionsQuintessence Int2015461917

12 

W Renne M Ludlow J Fryml Z Schurch A Mennito R Kessler Evaluation of the accuracy of 7 digital scanners: An in vitro analysis based on 3-dimensional comparisonsJ Prosthet Dent201711813642

13 

GJ Christensen Will Digital Impressions Eliminate the Current Problems With Conventional Impressions?J Am Dent Assoc200813967613

14 

GVS Salomão EP Chun RDS Panegaci FT Santos Analysis of Digital Workflow in ImplantologyCase Rep Dent20212021665590810.1155/2021/6655908

15 

SJ Lee SW Kim JJ Lee CW Cheong Comparison of Intraoral and Extraoral Digital Scanners: Evaluation of Surface Topography and PrecisionDent J (Basel)20208252



jats-html.xsl


This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.

  • Article highlights
  • Article tables
  • Article images

Article History

Received : 02-08-2023

Accepted : 28-08-2023